seca mVSA 535
Product Number 535 1030 002EMR-validated Spot-check Vital Signs Monitor
Order today, ships within 24 hours!
- Current mVSA configuration:
- Blood Pressure + Pulse
- Without SpO₂ module
- Genuis™ temperature modul
Quick Overview
Description
Technical Data
Precision
- Basic
- Advanced
- Premium
Key-Technical Data
Blood pressure measurement methodOscillometric, deflation or inflation measurement switchable between single measurement (seca EQ BP®) and configurable multiple measurements (BP 30)
Blood pressure measurement timeNormal: 15 - 20 sec / max. 90 sec (adults)
Temperature measurement response time< 2 sec.
Temperature measuring methodin Ear
Scope of Delivery
- 1x Genuis™ 2 Covidien Genius™ probe covers (Box á 96 pcs)
- Removable SmartBucket with integrated Vital Signs Moduls: for blood pressure/Temp/SpO2 measurement
- Genius™ 3 Tympanic in-Ear thermometer
- Monitor with power cable
- Blood pressure cuff (reusable), size M
Product Properties
- Easy vitals capture of blood pressure, temperature, pulse rate and SpO2 in less than 60 seconds
- Quick and accurate blood pressure measurement in under 20 seconds
- Configurable seca EasyQuick BP™ average, sprint or single measurement mode (EQ BP™) provides true screening of hypertension
- Clinical observations: Enter up to 24 different values like blood sugar, pain level or respiration rate and send them to the EMR
- seca scale connectivity through seca 360° proximity: One integration interface for seca EMR-validated devices
- Long lasting battery for up to 150 full sets of vitals on one single charge
- Convincing data storage: All measurements stored and available for later processing
- EMR-validated: transmit measured data directly to most Electronic Medical Record (EMR) systems
seca mVSA – Where ease meets accuracy
Acquiring patient’s vitals is a routine task, performed on almost every patient. The seca mVSA 535 was designed to make this task as easy and efficient as possible. Staff at any level can operate the spot-check monitor due to its intuitive touch screen and self-explanatory user interface. The vitals capture is simple and fully validated for HL7 and many EMR systems. Connectivity to seca scales and measuring stations due to seca 360° proximity is also a plus as well as the unique seca EQ BP™ technology. The seca mVSA fits perfectly into your existing workflows and frees up time and resources to improve patient care.
Precision
- Basic
- Advanced
- Premium
Key-Technical Data
Blood pressure measurement methodOscillometric, deflation or inflation measurement switchable between single measurement (seca EQ BP®) and configurable multiple measurements (BP 30)
Blood pressure measurement timeNormal: 15 - 20 sec / max. 90 sec (adults)
Temperature measurement response time< 2 sec.
Temperature measuring methodin Ear
Functions and Properties
Scope of Delivery
- 1x Genuis™ 2 Covidien Genius™ probe covers (Box á 96 pcs)
- Removable SmartBucket with integrated Vital Signs Moduls: for blood pressure/Temp/SpO2 measurement
- Genius™ 3 Tympanic in-Ear thermometer
- Monitor with power cable
- Blood pressure cuff (reusable), size M
Technical Data
Power supplyBuilt-in power adapter, internal rechargeable lithium-ion battery
Turn-off time with power supplyNo switch off
InterfacesEthernet
USB 2.0
Wi-Fi
USB 2.0
Wi-Fi
Display7" Touchscreen-Display
Net weight5.9 lbs
2.7 kg
2.7 kg
Product width9.9 inch
252 mm
252 mm
Product height10.3 inch
262 mm
262 mm
Product depth10.9 inch
278 mm
278 mm
Blood pressure measurement methodOscillometric, deflation or inflation measurement switchable between single measurement (seca EQ BP®) and configurable multiple measurements (BP 30)
Blood pressure measurement certaintyExcess pressure limit 300 mmHg;
automatic pressure release at > 300 mmHg
automatic pressure release at > 300 mmHg
Blood pressure measurement timeNormal: 15 - 20 sec / max. 90 sec (adults)
Measurement accuracy at ambient temperature: 16 °C – 33 °C (60.8 °F – 91.4 °F)Target temperature: 36.7 °C 38.9 °C (98.1 °F 102 °F) ± 0.1 °C ( ± 0.2 °F)
Measurement accuracy at target temperature: 25 °C (77 °F)Target temperature: 33 °C 42 °C (91.4 °F 107.6 °F) ± 0.2 °C ( ± 0.4 °F)
Blood pressure measurement range – inflationpSYS: 77 - 200 mmHg pDIA: 45 - 190 mmHg pMAP: 56 - 193 mmHg
Blood pressure measurement range – deflationpSYS: 25 - 280 mmHg pDIA: 10 - 220 mmHg pMAP: 15 - 260 mmHg
Blood pressure measurement accuracy (under laboratory conditions)Mean deviation pSYS, pDIA, pMAP: max. +/- 3 mmHg
Blood pressure measurement accuracy – inflation (clinically validated according to ISO 81060)Mean deviation systole 0.94 mmHg; Standard deviation systole 3.84 mmHg; Mean deviation diastole 0.57 mmHg; Standard deviation diastole 3.17 mmHg
Blood pressure measurement accuracy – deflation (clinically validated according to ISO 81060)Mean deviation systole 0.39 mmHg; Standard deviation systole 2.57 mmHg; Mean deviation diastole 0.43 mmHg; Standard deviation diastole 1.73 mmHg
Blood pressure pulse rate range – inflation45 - 200 bpm
Blood pressure pulse rate range – deflation30 - 240 bpm
Blood pressure pulse rate accuracymax. +/- 3 bpm
SpO₂ moduleWithout SpO₂ module
Temperature moduleGenuis™ temperature modul
Temperature measuring range33 - 42 °C / 91.4 - 107.6 °F
Temperature measurement response time< 2 sec.
Temperature measuring methodin Ear
Functions and Properties
- Blood pressure measurement
- Temperature measuement